Technical Note: Investigating of dosimetric leaf gap and leaf transmission factor variations across gantry and collimator angles in volumetric modulation arc therapy

Download (1.24 MB)
journal contribution
submitted on 2025-08-13, 11:28 and posted on 2025-08-13, 11:29 authored by Aram Rostami, Mojtaba Barzegar, Muhammad Usman, Satheesh Prasad Paloor, Abbass Yousef Mkanna, Alla Fuad Al‐Sabahi, Rabih Wafiq Hammoud
<h3>Purpose</h3><p dir="ltr">This study investigates the influence of gantry and collimator angles on the dosimetric leaf gap (DLG) and leaf transmission factor (LTF) in a Varian LINAC equipped with rounded‐end multi‐leaf collimators (MLCs). While Varian guidelines recommend DLG measurements at zero degrees for both gantry and collimator, this research aims to address the knowledge gap by assessing DLG and LTF variations at different gantry and collimator angles.</p><h3>Methods</h3><p dir="ltr">Measurements were conducted using a Varian TrueBeam LINAC with a Millennium 120‐leaf MLC and Eclipse TPS version 16.1. The beams utilized in this study had energies of 6 MV, 10 MV, 6 FFF, and 10 FFF. LTF and DLG were determined using ionization chambers in solid water phantoms at various gantry angles (0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°). For each gantry angle, measurements were also taken at various collimator angles (0°, 45°, 90°, and 315°). Dosimetric impacts were evaluated through VMAT Picket Fence tests and patient‐specific verification using portal dosimetry for 10 clinical VMAT plans.</p><h3>Results</h3><p dir="ltr">LTF values showed no significant variation across gantry and collimator angles. However, DLG values exhibited notable differences depending on the gantry angle and were independent of the collimator angle. The highest DLG value was observed at a gantry angle of 270 degrees, while the lowest was at 90 degrees. The AXB DLGAverage (averaging seven measurements of DLGs at different gantry angles) model demonstrated the best agreement between measured and calculated dose distributions, indicating the importance of considering averaged DLG values across multiple gantry angles for accurate dose calculations.</p><h3>Conclusion</h3><p dir="ltr">Our study highlights the variability of DLG with gantry angle alterations, contrary to Varian guidelines recommending DLG measurements at zero gantry angle only. We advocate for utilizing an averaged DLG value from measurements across multiple gantry angles, as outlined in our methodology.</p><h2>Other Information</h2><p dir="ltr">Published in: Journal of Applied Clinical Medical Physics<br>License: <a href="http://creativecommons.org/licenses/by/4.0/" target="_blank">http://creativecommons.org/licenses/by/4.0/</a><br>See article on publisher's website: <a href="https://dx.doi.org/10.1002/acm2.14523" target="_blank">https://dx.doi.org/10.1002/acm2.14523</a></p>

Funding

Open Access funding provided by the Qatar National Library.

History

Language

  • English

Publisher

Wiley

Publication Year

  • 2024

License statement

This Item is licensed under the Creative Commons Attribution 4.0 International License.

Institution affiliated with

  • Hamad Medical Corporation
  • National Center for Cancer Care and Research - HMC

Usage metrics

    National Center for Cancer Care and Research - HMC

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC