Digital radiography image quality evaluation using various phantoms and software

Download (4.39 MB)
journal contribution
submitted on 2025-05-19, 07:10 and posted on 2025-05-19, 07:11 authored by Ioannis A. Tsalafoutas, Shady AlKhazzam, Virginia Tsapaki, Huda AlNaemi, Mohammed Hassan Kharita

Purpose

To investigate the effect of the exposure parameters on image quality (IQ) metrics of phantom images, obtained automatically using software or from visual evaluation.

Methods

Three commercial phantoms and a homemade phantom constructed according to the instructions given in the IAEA Human Health Series No. 39 publication were used, along with the respective software that estimate automatically various IQ metrics. Images with various exposure parameters were acquired in a digital radiography (DR) unit. For the commercial phantoms, visual evaluations were also performed. The IQ scores obtained were analyzed to investigate the effects of increasing incident air kerma (IAK), tube potential (kVp), additional filtration, and acquisition protocol on IQ.

Results

The effects of the exposure parameters on the IQ metrics, determined with the commercial and the IAEA phantoms, were not the same. For example, clear trends of improvement of IQ scores with increased IAK and reduction of most IQ scores with increased kVp were observed mostly with the IAEA phantom, but not with the commercial phantoms (for both automatic and visual scoring methods). For all phantoms, the maximum variations in IQ scores observed for repeated identical exposures were almost always below 10% with automatic evaluation whereas, for visual evaluation, reached 17%.

Conclusions

Failure to detect some expected trends with the complex commercial phantoms may be attributed to the fact that IQ in DR is more strongly affected by the post‐processing procedures, which may mask the effect of other parameters on IQ, something that was not observed with the simple IAEA phantom.

Other Information

Published in: Journal of Applied Clinical Medical Physics
License: http://creativecommons.org/licenses/by/4.0/
See article on publisher's website: https://dx.doi.org/10.1002/acm2.13823

Funding

Open Access funding provided by the Qatar National Library.

History

Language

  • English

Publisher

Wiley

Publication Year

  • 2022

License statement

This Item is licensed under the Creative Commons Attribution 4.0 International License.

Institution affiliated with

  • Hamad Medical Corporation
  • Weill Cornell Medicine - Qatar

Usage metrics

    Hamad Medical Corporation

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC